Skip to product information
1 of 1

ValleyLab Force 1C Electrosurgery Unit w/ Foot Pedal - Reconditioned

ValleyLab Force 1C Electrosurgery Unit w/ Foot Pedal - Reconditioned

Regular price $2,195.00
Regular price Sale price $2,195.00
Sale Sold out

Valleylab Force 1C Electrosurgical Generator w/ Monopolar Foot Switch

 

OPTIONS:

  • Bipolar foot switch and
  • Rolling cart.

Standard Accessories Available

  • Handpieces
  • Grounding Pads

 

FEATURES:

 

The Force 1C has a streamlined, compact design perfect for the requirements of outpatient surgery - as well as the power and performance features needed for a full range of general inpatient surgical procedures.

THREE BLENDED CUT MODES provide flexibility with varying degrees of hemostasis in the cut modes.

QUICK SET UP FEATURES reduce the need for readjustment between procedures. Microbipolar and monopolar modes are set independently, and the digital display allows easy confirmation of power levels.

ISOLATED OUTPUT DESIGN directs…

  • Height: 6 in. (15 cm)
  • Width: 11 in. (28 cm)
  • Length: 17 in. (43 cm)
  • Weight: 15 lbs. (6.8kg)

OUTPUT WAVEFORMS
  • Cut: 516 kHz sinusoid
  • Blend 1: 516 kHz sinusoid bursts at 70% duty cycle recurring at 31 kHz
  • Blend 2: 516 kHz sinusoid bursts at 45% duty cycle recurring at 31 kHz
  • Blend 3: 516 kHz sinusoid bursts at 20% duty cycle recurring at 31 kHz
  • Coag: 516 kHz sinusoid bursts with a repetition frequency of 31 kHz
  • Microbipolar: 516 kHz sinusoid

Output power changes by less than 5% as the line voltage varies from 85-135 volts

LOW FREQUENCY LEAKAGE (50/60 HZ)
Source current, patient leads, all outputs tied together

  • Normal polarity, intact chassis ground < 10 µA
  • Normal polarity, ground open < 100 µA
  • Reverse polarity, ground open < 100 µA
  • Sink current, 135V applied, all inputs < 150 µA

HIGH FREQUENCY LEAKAGE
Less than 150 mA rms

INPUT POWER REQUIREMENST
Operation range is 85 to 135 AC volts. Current is less than 5 ampers in cut and less than 3 amperes in coag.

POWER READOUTS
Agree with actual power into rated load to within ±20% or 5 watts, whichever is greater.

REM SYSTEM
Measurement Frequency: 140 kHz ±20 kHz

Measurement Current: 3 mA maximum

Acceptable Resistance Ranges:
  • REM pad --5-135 ohms
  • Non-REM pad --less than 20 ohms

COOLING
Convection, no fan

AUDIO VOLUME
The mode indicator tones are adjustable from 45 to 65 dBa
The alarm tones are not adjustable and are set to 65 dBa
Designed to meet UL and CSA specifications.

 

 

View full details
Your cart
Product Product subtotal Quantity Price Product subtotal
ValleyLab Force 1C Electrosurgery Unit w/ Foot Pedal - Reconditioned
ValleyLab Force 1C Electrosurgery Unit w/ Foot Pedal - ReconditionedVL-F1C
ValleyLab Force 1C Electrosurgery Unit w/ Foot Pedal - ReconditionedVL-F1C
$2,195.00/ea
$0.00
$2,195.00/ea $0.00