Other products by ValleyLab

Description  more details

Valleylab Force 1C Electrosurgical Generator w/ Monopolar Foot Switch

 

OPTIONS:

  • Bipolar foot switch and
  • Rolling cart.

Standard Accessories Available

  • Handpieces
  • Grounding Pads

 

FEATURES:

 

The Force 1C has a streamlined, compact design perfect for the requirements of outpatient surgery - as well as the power and performance features needed for a full range of general inpatient surgical procedures.

THREE BLENDED CUT MODES provide flexibility with varying degrees of hemostasis in the cut modes.

QUICK SET UP FEATURES reduce the need for readjustment between procedures. Microbipolar and monopolar modes are set independently, and the digital display allows easy confirmation of power levels.

ISOLATED OUTPUT DESIGN directs…

  • Height: 6 in. (15 cm)
  • Width: 11 in. (28 cm)
  • Length: 17 in. (43 cm)
  • Weight: 15 lbs. (6.8kg)

OUTPUT WAVEFORMS
  • Cut: 516 kHz sinusoid
  • Blend 1: 516 kHz sinusoid bursts at 70% duty cycle recurring at 31 kHz
  • Blend 2: 516 kHz sinusoid bursts at 45% duty cycle recurring at 31 kHz
  • Blend 3: 516 kHz sinusoid bursts at 20% duty cycle recurring at 31 kHz
  • Coag: 516 kHz sinusoid bursts with a repetition frequency of 31 kHz
  • Microbipolar: 516 kHz sinusoid

Output power changes by less than 5% as the line voltage varies from 85-135 volts

LOW FREQUENCY LEAKAGE (50/60 HZ)
Source current, patient leads, all outputs tied together

  • Normal polarity, intact chassis ground < 10 µA
  • Normal polarity, ground open < 100 µA
  • Reverse polarity, ground open < 100 µA
  • Sink current, 135V applied, all inputs < 150 µA

HIGH FREQUENCY LEAKAGE
Less than 150 mA rms

INPUT POWER REQUIREMENST
Operation range is 85 to 135 AC volts. Current is less than 5 ampers in cut and less than 3 amperes in coag.

POWER READOUTS
Agree with actual power into rated load to within ±20% or 5 watts, whichever is greater.

REM SYSTEM
Measurement Frequency: 140 kHz ±20 kHz

Measurement Current: 3 mA maximum

Acceptable Resistance Ranges:
  • REM pad --5-135 ohms
  • Non-REM pad --less than 20 ohms

COOLING
Convection, no fan

AUDIO VOLUME
The mode indicator tones are adjustable from 45 to 65 dBa
The alarm tones are not adjustable and are set to 65 dBa
Designed to meet UL and CSA specifications.

 

 

 

Specifications:
Dimensions: 6 x 11 x 17 inches (15 x 28 x 43 cm)
Weight: 15 lbs. (6.8 kg)
 

Output Waveform:

Cut: 516 kHz sinusoid; Blend 1:516 kHz of sinusoid at 70% duty cycle recurring at 31 kHz; Blend 2: 516 kHz bursts of sinusoid at 45% duty cycle recurring at 31 kHz; Blend 3: 516 kHz bursts of sinusoid at 20% duty cycle recurring at 31 kHz; Coag: 516 kHz damped sinusoidal bursts with a repetition frequency of 31 kHz; Microbipolar: 516 kHz sinusoid, unmodulated.
 

Output Characteristics:    Max. (open circuit)           Rated Load             Nominal Power             Crest Factor
                                                P-P voltage                       ohms                        at rated load watts        at rated load typical
Cut                                         2400                                   300                            200                                 1.8 @100 W
Blend 1                                  2800                                   300                           175                                  2.7 @100 W
Blend 2                                  3200                                   300                            150                                 3.3 @100 W
Blend 3                                  3600                                   300                            125                                 4.6 @100 W
Coag                                       5200                                  300                            75                                    8.0 @50 W
Microbipolar                          900                                    100                              50                                   1.8 @40 W
 

Input Power source:

Max. operating range: 85-135 or 170-270 VAC; Nominal operating range: 110-120 or 220-240 VAC
Max. Current                                                           Max. Power
Idle 0.5 A                                                                          60 W
Cut 5.0 A                                                                          600 W
Coag 3.0 A                                                                      400 W
Microbipolar 2.0 A                                                          180 W
The line frequency may vary between 50 and 60 Hz

Line Regulation:

Between 85-135 volts or 170-270 volts input, output power into nominal load will vary no more than 20% or 5 watts, whichever is greater
REM Contact Quality Mon.:

Measurement frequency: 140 kHz ± 20 kHz. Measurement current: 3.0 mA, maximum
Low Frequency Leakage:

Source current nominal polarity, intact chassis ground: <10 μA, ground open: <50 μA.

Source current reverse polarity, ground open: <50 μA. sink current, 135 VAC applied, all inputs: <20 μA.

Chassis source current, ground open: <100 μA
High Frequency Risk Parameters:

Microbipolar RF leakage current: <150 mA. Monopolar RF leakage current: < 150 mA
 

Email A Friend

Send your friend a link to this product.
 
 
 

Subscribe To Email List

Enter your email address below to subscribe to All customers.